Medical-Air-Desiccant-Dryers
  • Each desiccant dryer is individually sized for NFPA peak calculated demand.
  • Each dryer is capable of producing a 14°F (-10°C) pressure dew point at 100 psig inlet and 50 psig delivery.
  • Purge flow is minimized through an integral purge saving control system.
  • The desiccant is contained within a packed bed canister to prevent excessive downstream dusting.
  • 441™ transfer valve per dryer with ceramic plate design to acheive tower changeover
  • Easy alternation between on line/off line dryers by means of a valved bypass
  • The dryer will use up to 15% of the system capacity to purge
  • Maximum inlet temperature for dryers is 110°F
  • Anodized aluminum blocks with o-ring seals and SAE o-ring fitings to eliminate leak points